首頁 > 產品展示 > > 其他檢測儀器 > TC-500 氧氮聯(lián)合測定儀

TC-500 氧氮聯(lián)合測定儀

簡要描述:簡單說明:TC500系列儀器質量、性能、工藝滿足您對金屬、難熔物、其他無機材料日常生產中氮/氧分析的需要

  • 產品型號:TC-500
  • 廠商性質:生產廠家
  • 更新時間:2025-07-17
  • 訪  問  量:2560
詳情介紹

詳細說明:

TC500系列儀器質量、性能、工藝滿足您對金屬、難熔物、其他無機材料日常生產中氮/氧分析的需要

Determination of Oxygen and Nitrogen in
Reactive/Refractory Metals and Their Alloys*
 
Approval
ASTM E-1409 Oxygen Determination in Titanium and Titanium Alloys
ASTM E-1569 Oxygen Determination in Tantalum
ASTM E-1937 Nitrogen Determination in Titanium and Titanium Alloys
 
Sample Preparation
A clean representative sample is required in order to obtain optimum results. Solid samples can be leached in suitable acid or abraded with a clean file, rinsed in acetone, and dried with warm air. Refer to ASTM methods above for further details.
 
Accessories
782-720 Crucible; 782-721 Electrode Tip; 502-344 UHP Nickel
Baskets; 501-073 Graphite; 503-032 Glass Accelerator Scoop;
501-059 Tin Capsules—for powdered samples. (Additional steps required for powdered samples are noted in parenthesis.)
 
Calibration Standard
LECO 501-320 Titanium Pin, 502-047 Zirconium Pin, NIST, or other suitable reactive/refractory metal standard.
 
Method Parameters
Analysis Parameters
Outgas Cycles                           2
Analysis Delay (seconds)             20
Analysis Comparator                 1
Analysis Type                              Auto
 
Element Parameters                  Oxygen           Nitrogen
Minimum Time (seconds)
35
55
Integration Delay (seconds)
5
15
Comparator Level (%)
1
1
 
Furnace Parameters
Furnace Control Mode             Power
Purge Time (seconds)                    15
Outgas Time (seconds)                20
Cool Time (seconds)                    5
Outgas Power (Watts)                 6300
Analyze Power (Watts)                5300
 
 
 
 
 


 

*This includes Ti, Zr, W, Mo, Ta, Nb, Hf, and their alloys.

 
TC500


 

 
Typical Results
 
 
 
Titanium Pin
Mass (g)
Oxygen (%)
Nitrogen (%)
 
0.1135
0.1808
0.0189
 
0.1133
0.1843
0.0194
 
0.1134
0.1831
0.0184
 
0.1138
0.1819
0.0188
 
0.1133
0.1821
0.0195
 
0.1131
0.1829
0.0185
 
0.1134
0.1833
0.0191
 
0.1143
0.1849
0.0194
 
0.1125
0.1837
0.0189
 
0.1072
0.1830
0.0191
 
Average
0.1830
0.0190
 
Std. Dev.
0.0012
0.0004
 
Zirconium Wire
 
Mass (g)
 
Oxygen (%)
 
Nitrogen (%)
 
0.1067
0.1246
0.0017
 
0.0978
0.1273
0.0017
 
0.1180
0.1246
0.0019
 
0.1097
0.1235
0.0014
 
0.1141
0.1272
0.0017
 
0.1017
0.1226
0.0018
 
0.1140
0.1269
0.0020
 
0.0987
0.1239
0.0020
 
0.1015
0.1233
0.0022
 
0.1197
0.1270
0.0020
 
Average
0.1251
0.0018
 
Std. Dev.
0.0018
0.0002
 
Procedure
 
 
 
1.   Determine the blank as follows:
a.   Enter the "blank" ID code with a 1.0000 g weight in the weight stack.
b.   Press the loader control switch, the sample loader will open.
c.   Place one 502-344 UHP Nickel basket into the loading head using clean tweezers.
(Place a 501-059 Tin capsule into the Nickel basket before placing it in the loading head.)
d.   Press the loader control switch, the sample loader will close and seal and the furnace electrode will open.
e.   Remove crucible from electrode tip and discard.  Clean furnace area using the appropriate brushes.  Vacuum away loose dust.
f.    Place ~0.05 g 501-073 into the bottom of a 782-720 crucible.  ~0.05 g is approximay a ¼ full
503-032 glass accelerator scoop.
g.   Place the crucible on the lower electrode.
h.   Press the loader control switch, the furnace electrode will close and the analysis sequence will start automatically.
i.    Repeat steps 1a through 1h at least four more times.
j.    Enter blank following routine outlined in operator's instruction manual.
2.   Calibrate the instrument as follows:
a.   Weigh the calibration sample. (Weigh the calibration sample into the tin capsule.)
b.   Enter the calibration sample ID code and sample weight in the weight stack.
c.   Place the calibration sample (capsule) into a nickel basket.
d.   Press the loader control switch, the sample loader will open.
e.   Carefully place the calibration sample/nickel basket (capsule) into the loading head using clean tweezers.  Make sure that the calibration sample (capsule) stays in the basket and the basket
stays upright.
f.    Press the loader control switch, the sample loader will close and seal and the furnace electrode will open.
continued on page 3


 

Procedure (continued from page 2)
g.   Remove crucible from electrode tip and discard.  Clean furnace area using the appropriate brushes.  Vacuum away loose dust.
h.   Place ~0.05 g graphite into the bottom of a crucible.
i.    Place the crucible on the lower electrode.
j.    Press the loader control switch, the furnace electrode will close and the analysis sequence will start automatically.
k.   Repeat steps 2a through 2j a minimum of three times.
l.    Complete a calibration by following the auto calibration procedure as outlined in the operator's instruction manual.
m.  Verify the calibration by analyzing the calibration sample again.  It should fall within the expected tolerances.  If not, repeat steps 2a through 2l again.
3.   Analyze the samples as follows:
a.   Weigh ~0.1 g sample. (Weigh the sample into the tin capsule.)
b.   Enter the sample ID code and sample weight in the weight stack.
c.   Place the sample (capsule) into a nickel basket.
d.   Press the loader control switch, the sample loader will open.
e.   Carefully place the sample/nickel basket (capsule) into the loading head using clean tweezers.
Make sure that the sample (capsule) stays in the basket and the basket stays upright.
f.    Press the loader control switch, the sample loader will close and seal and the furnace electrode will open.
g.   Remove crucible from electrode tip and discard.  Clean furnace area using the appropriate brushes.  Vacuum away loose dust.
h.   Place ~0.05 g graphite into the bottom of a crucible.
i.    Place the crucible on the lower electrode.
j.    Press the loader control switch, the furnace electrode will close and the analysis sequence will start automatically.
 
Theory of Operation
The TC500 is a microprocessor-based, software-controlled instrument that measures both nitrogen and oxygen in a wide variety of metals, refractories, and inorganic materials. The inert gas fusion principle is employed. A weighed sample, placed in a high-purity graphite crucible, is fused under a flowing helium gas stream at temperatures sufficient to release oxygen, nitrogen, and hydrogen. The oxygen in the
sample, combines with the carbon from the crucible to form carbon monoxide. The nitrogen present in
the sample releases as molecular nitrogen, and any hydrogen present is released as hydrogen gas. The helium carries the sample gases through heated rare earth copper oxide which converts carbon
monoxide to carbon dioxide (CO2) and hydrogen to water (H2O). The nitrogen passes through
unchanged. The gases are then passed through a CO2  infrared (IR) cell where the oxygen is measured as
CO2. CO2  and H2O are then removed by a Lecosorb/Anhydrone trap, while the nitrogen passes through
to a thermal conductivity (TC) cell for determination.


 

TC500 Flow Diagram
 
 
 
 
 

留言框

  • 產品:

  • 您的單位:

  • 您的姓名:

  • 聯(lián)系電話:

  • 常用郵箱:

  • 省份:

  • 詳細地址:

  • 補充說明:

  • 驗證碼:

    請輸入計算結果(填寫阿拉伯數(shù)字),如:三加四=7
技術支持:化工儀器網   sitemap.xml   管理登陸
©2025 版權所有:寧波市鄞州瑾瑞儀器設備有限公司   備案號:浙ICP備14028217號-3

浙公網安備33020402000311號

娇妻被交换黑人粗又大又硬| 中文字幕久精品视频在线| 蜜臀国产av一区二区三区| 欧洲精欧洲精品高清一区| 超短裙流水白浆在线播放| 日本一欧美一欧美一亚洲| 精品偷自拍另类在线观看| 久久国产劲暴∨内射新川| 差差漫画页面画在线阅读弹窗| 久久精品国产第一区二区| jizzjizz欧美69巨大| 日本亚洲欧美男人的天堂| 国产精品自在拍在线播放| 国产v精品欧美精品v日韩| 中文在线中文资源| 插入小穴黄视频在线观看| 美女全黄网站在线免费看| 国产成人综合亚洲精品国产| 性色XXXXHD天美传媒| 中文字幕亚洲制服在线看| 最新午夜男女福利片视频| 久久综合l88熟人妻3| 亚洲无码一级片在线播放| 啊灬啊灬啊灬快灬高潮了听书| 欧美精品一二三区在线视频| 曰曰摸日日碰夜夜爽歪歪| 人妻无码熟妇乱又伦精品| 亚洲日韩国产一区二区 | 欧美va亚洲va日韩va| 久久国产精品无码网站| 日本16岁RAPPER| 成人三级在线观看一区二区| 夫の前で痴汉に绝顶人妻国产剧 | 久久久国产99精品视频| 2022最新韩国理伦片在线观看| 国产欧美亚洲精品a| 欧美一区二区aa大片蜜桃| 久青草香蕉精品视频在线| 中文av人妻av无码中文| 99线手机在线观看视频| 一区二区三区自拍偷拍av|